Echocardiography & valvular
Prosthetic valve assessment
Grades aortic and mitral prosthesis Doppler against ASE/EACVI criteria and calculates indexed EOA for patient–prosthesis mismatch. Sets the ESC and ACC/AHA antithrombotic plans for mechanical, surgical bioprosthetic and TAVI valves side by side, including INR targets by prosthesis thrombogenicity, aspirin, DOAC contraindications and valve-thrombosis management. ESC tags refer to the 2021 ESC/EACTS guideline; check them against the 2025 update.
Upload a report, lab sheet or echo picture to fill this calculator automatically.
Prosthesis
Prosthesis thrombogenicity (ESC)
Doppler
Aortic prosthesis criteria (ASE / EACVI). Enter what you measured
Normal
< 3
Possible
3–4
Significant
> 4
Normal
< 20
Possible
20–35
Significant
> 35
Normal
≥ 0.30
Possible
0.25–0.29
Significant
< 0.25
Normal
> 1.2
Possible
0.8–1.2
Significant
< 0.8
Normal
< 80
Possible
80–100
Significant
> 100
Patient factors
Set INR targets and the need for anticoagulation
Aortic mechanical · 0 Doppler parameters
Antithrombotic plan
Risk factor for INR target: no
Antithrombotic therapy
ESC
- Lifelong VKA, target INR 2.5Class IESC 2021Median target set by prosthesis thrombogenicity and patient risk factors (mitral/tricuspid position, prior thromboembolism, AF, mitral stenosis, LVEF < 35%). INR self-management recommended in trained patients.
- DOACs contraindicatedIIIESC 2021Excess thrombosis and bleeding with dabigatran (RE-ALIGN) and apixaban with On-X (PROACT Xa).
ACC/AHA
- Lifelong VKA, target INR 2.5Class IACC/AHA 2020Bileaflet or current-generation tilting-disc aortic valve without risk factors.
- Add aspirin 75–100 mgIIbACC/AHA 2020May be considered when bleeding risk is low (downgraded from routine use in 2014).
- DOACs contraindicatedIIIACC/AHA 2020Excess thrombosis and bleeding with dabigatran (RE-ALIGN) and apixaban with On-X (PROACT Xa).
Valve thrombosis
ESC
- Obstructive mechanical valve thrombosisClass IESC 2021Urgent or emergency surgery. Fibrinolysis only if surgery is not available or very high risk, or for right-sided valves.
ACC/AHA
- Symptomatic mechanical valve thrombosisClass IACC/AHA 2020Either slow-infusion low-dose fibrinolysis or emergency surgery, chosen by the Heart Team on thrombus size, NYHA class, surgical risk and local expertise.
Get a baseline TTE 6 weeks to 3 months after implantation. Later studies are judged against it.
References
- Zoghbi WA, et al. Recommendations for evaluation of prosthetic valves with echocardiography and Doppler ultrasound (ASE). J Am Soc Echocardiogr 2009;22:975-1014.
- Lancellotti P, et al. Recommendations for the imaging assessment of prosthetic heart valves (EACVI). Eur Heart J Cardiovasc Imaging 2016;17:589-590.
- VARC-3 Writing Committee. Valve Academic Research Consortium 3: updated endpoint definitions for aortic valve clinical research. Eur Heart J 2021;42:1825-1857.
- Vahanian A, et al. 2021 ESC/EACTS Guidelines for the management of valvular heart disease. Eur Heart J 2022;43:561-632.
- Otto CM, et al. 2020 ACC/AHA Guideline for the management of patients with valvular heart disease. Circulation 2021;143:e72-e227.
- Praz F, et al. 2025 ESC/EACTS Guidelines for the management of valvular heart disease. Eur Heart J 2025.
More in Echocardiography & valvular
- LV systolic function — grade
- LV diastolic function — echo grading
- LV hypertrophy — mass & geometry
- Aortic stenosis — echo grading
- Aortic regurgitation — echo grading
- Mitral regurgitation — echo grading
- Mitral stenosis — echo grading
- Wilkins Score
- Right ventricle, IVC & PH probability
- IVC — collapsibility & RA pressure